Tt. Ioannidis et al., LONG-TERM BEHAVIOR OF THE CHARNLEY OFFSET-BORE ACETABULAR CUP, Journal of bone and joint surgery. British volume, 80B(1), 1998, pp. 48-53
We report the long-term radiological results of 58 total hip arthropla
sties (THA) using the Charnley offset-bore acetabular socket. Wear was
measured at four sites and radiolucent lines and possible migration w
ere recorded. Four cups were retrieved at revision and were examined u
sing light microscopy, SEM and X-ray microanalysis. At a mean follow-u
p of seven years the mean wear in the DeLee and Charnley zone I was 0.
4 mm and in zone II 0.26 mm. The wear rate was 0.06 and 0.04 mm/year,
respectively. Progression of radiolucent lines was seen in five cases
(8.6%). Three sockets (5.2%) were revised because of aseptic migration
at a mean follow-up of 9.8 years and one socket for infection at two
years. The offset-bore acetabular cup had excellent wear behaviour and
a low migration and revision rate. We recommend that it should be con
sidered in THA since the use of small cups is increasing, particularly
in revision cases.