USE OF BUBBLE DETECTORS TO MEASURE NEUTRON CONTAMINATION OF A MEDICALACCELERATOR PHOTON-BEAM

Citation
L. Bourgois et al., USE OF BUBBLE DETECTORS TO MEASURE NEUTRON CONTAMINATION OF A MEDICALACCELERATOR PHOTON-BEAM, Radiation protection dosimetry, 74(4), 1997, pp. 239-246
Citations number
14
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
74
Issue
4
Year of publication
1997
Pages
239 - 246
Database
ISI
SICI code
0144-8420(1997)74:4<239:UOBDTM>2.0.ZU;2-S
Abstract
Medical electron accelerators generate spurious neutron radiations: in order to protect patients, neutron radiation levels must remain below internationally recommended values. In the present study, bubble dete ctors were used to measure the neutron component in the primary beam a nd outside the radiation field for 8,15, 18 and 25 MV photons from a S aturne 43F type medical accelerator. The results are compared with tho se obtained using other methods. Values of between 0.25 and 7.5 mSv.Gy (-1) were determined at the beam isocentre for photons of 8 to 25 MV; these values are in agreement with those published in the literature.