Cf. Wang et al., RAPID-DETERMINATION OF SR-90 BY COMPTON SUPPRESSION GAMMA-RAY SPECTROMETRY, Journal of radioanalytical and nuclear chemistry, 223(1-2), 1997, pp. 157-162
This work presents a low-level background gamma spectrometry system ca
pable of achieving rapid determination of Sr-90 without chemical separ
ation, in a soil sample when the Sr-90 to gamma-ray emitting nuclide c
oncentration ratio exceeds 10. The system consists of a highly efficie
nt HPGe central detector operated in coincidence or anticoincidence wi
th a NaI(Tl) guard detector and a NaI(Tl) backscatter detector Experim
ental results indicate that the bremsstrahlung radiation from a pure b
eta-emitter can be distinguished with respect to the differences in no
rmal and coincidence spectra. The Sr-90 activity can be directly deter
mined in the sample in less than 1 hour with a minimum detectable conc
entration (MDC) estimated at 0.002 Bq . g(-1) for a sample containing
no appreciable gamma-ray nuclides or other high energy, beta-emitting
nuclides. In actual measurement, a 0.06 Bq . g(-1) MDC for contaminate
d soils is achieved.