Bv. Darden et al., A COMPARISON OF IMPEDANCE AND ELECTROMYOGRAM MEASUREMENTS IN DETECTING THE PRESENCE OF PEDICLE WALL BREAKTHROUGH, Spine (Philadelphia, Pa. 1976), 23(2), 1998, pp. 256-262
Study Design. A prospective comparison of impedance measurements, elec
trically elicited electromyograms, and mechanically elicited electromy
ograms to detect pedicle well breakthrough. Objective. To determine wh
ether impedance measurements are as Sensitive as electromyogram measur
ements in evaluating pedicle well breakthrough. Summary of Background
Data. In a previous animal study, impedance values in pedicle screw pl
acement were tested, to determine a baseline value for an intact pedic
le. If pedicle wall breakthrough occurred, it was thought that the imp
edance values should be significantly lower. Methods. Impedance measur
ements, electrically elicited electromyograms and mechanically elicite
d electromyograms were recorded in 20 patients undergoing surgery for
spinal degeneration, using previously de-scribed standard protocol. An
alysis of variance statistics were used to evaluate the data. Results.
Impedance values for the pedicle holes varied from 500 ohms to 31,000
ohms. There was no correlation between these values and those of the
two pedicles in which breaches were detected on visual inspection. Ele
ctrically elicited electromyograms detected the breakthroughs in both
pedicles, mechanically elicited electromyograms detected one of the br
eakthroughs. Conclusions. Electrically elicited electromyograms were m
ore sensitive in detecting pedicle wall breakthrough than were impedan
ce measurements. This may be because of the inability to ascribe absol
ute impedance values to human pedicle bone caused by the wide variabil
ity in bone quality.