SLURRY SAMPLING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY FOR TRACE-ELEMENT ANALYSIS OF HIGH-PURITY TUNGSTEN TRIOXIDE

Citation
V. Krivan et al., SLURRY SAMPLING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY FOR TRACE-ELEMENT ANALYSIS OF HIGH-PURITY TUNGSTEN TRIOXIDE, Fresenius' journal of analytical chemistry, 360(2), 1998, pp. 167-174
Citations number
33
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
360
Issue
2
Year of publication
1998
Pages
167 - 174
Database
ISI
SICI code
0937-0633(1998)360:2<167:SSEASF>2.0.ZU;2-K
Abstract
A slurry sampling ETAAS method for the determination of trace concentr ations of Ca, Co, Cr, Cu, Fe, K, Mg, Na and Ni in tungsten trioxide is presented. A high background absorption appearing with large sample a mounts during atomization at high temperatures was the only matrix int erference observed. It could be significantly reduced by volatilizatio n of the matrix in the pyrolysis step. Calibration curves recorded wit h aqueous standards could be used for calibration in all cases excludi ng the de termination of very low analyte concentrations of Co and Fe which required the use of the standard additions method. The results o btained by this method showed an excellent agreement with those determ ined in digests by ETAAS. The detection limits achievable by the slurr y technique were between 1 (Ca, K, ME, Na) and 210 (Cu) ng/g. They wer e mostly by more than one order of magnitude better than those of the solution techniques.