We introduce a new design for the optical cuvette and a nem optical la
y-out for the Scanning Flow Cytometer (SFC) that permits measurement o
f the angular dependency of the scattered Light from individual moving
particles, The improved optical scheme of the SFC allows measurement
of the angular scattering pattern of individual particles at polar ang
les from 10 degrees to 120 degrees with integration at azimuthal angle
s from 0 degrees to 360 degrees and with angular resolution of better
than 0.5 degrees. The performance of the SFC is demonstrated using cer
tified polystyrene particles as reference material, The aim of this wo
rk is to develop a flow cytometer, which, by recording the entire ligh
t scattering pattern of individual biological particles, would provide
more information about the particle structure than the ordinary wide
angle, forward and side scattering concepts. (C) 1998 Wiley-Liss, Inc.