GSO-DETECTOR SYSTEM NYMPHS FOR (LI-7, BE-7-GAMMA) REACTION AT INTERMEDIATE ENERGIES

Citation
S. Nakayama et al., GSO-DETECTOR SYSTEM NYMPHS FOR (LI-7, BE-7-GAMMA) REACTION AT INTERMEDIATE ENERGIES, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 404(1), 1998, pp. 34-40
Citations number
8
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
404
Issue
1
Year of publication
1998
Pages
34 - 40
Database
ISI
SICI code
0168-9002(1998)404:1<34:GSNF(B>2.0.ZU;2-V
Abstract
A gamma-ray detector system ''NYMPHS'' was built for Be-7-gamma coinci dence measurements to separate (Li-7, Be-7(g.s)) and (Li-7, B-7(0.43Me V)) channels in the (Li-7,Be-7) reaction at intermediate energies. NYM PHS is composed of 22 GSO scintillators, each with a volume of about 3 0 cm(3). A typical energy resolution was about 9% at E-gamma = 0.66 Me V. The absolute detection efficiency of NYMPHS was 0.200 +/- 0.005 for the 0.43 MeV Be-7 gamma-ray. NYMPHS combined with a magnetic spectrog raph showed good performance for Be-7-gamma coincidence measurements i n the (Li-7,Be-7) reactions on C-12 and Zr-90 targets at E-L = 65A MeV . (C) 1998 Elsevier Science B.V. All rights reserved.