Gl. Driscoll et al., VARIATION IN THE DETERMINATION OF FOLLICULAR DIAMETER - AN INTER-UNITPILOT-STUDY USING AN ULTRASONIC PHANTOM, Human reproduction, 12(11), 1997, pp. 2465-2468
Ultrasound operators in assisted reproductive technology units in New
South Wales, Australia cooperated in a study that was conducted to ass
ess their ability to measure fixed objects embedded in an ultrasound p
hantom. The results have demonstrated a large variation ranging betwee
n 10 and 25% coefficient of variation on distances between 10 and 32 m
m. As ultrasonic imaging is the only component of assisted reproductiv
e technology not currently controlled by an external quality assurance
scheme, the authors suggest there may be a need to establish a progra
mme.