VALIDATION OF AN IN-OFFICE DENTAL UNIT WATER MONITORING TECHNIQUE

Citation
Ri. Karpay et al., VALIDATION OF AN IN-OFFICE DENTAL UNIT WATER MONITORING TECHNIQUE, The Journal of the American Dental Association, 129(2), 1998, pp. 207-211
Citations number
29
Categorie Soggetti
Dentistry,Oral Surgery & Medicine
ISSN journal
00028177
Volume
129
Issue
2
Year of publication
1998
Pages
207 - 211
Database
ISI
SICI code
0002-8177(1998)129:2<207:VOAIDU>2.0.ZU;2-F
Abstract
The authors conducted a study to determine the sensitivity, specificit y and accuracy of HPC Samplers (Millipore Corp.) as compared with that of R2A agar and HPC agar when used to monitor dental unit water, or D UW, quality. Results were either a pass (200 colony-forming units per milliliter, or CFU/mL, or less) or a fail (greater than 200 CFU/mL). O f the 408 tests conducted, the results of the HPC Sampler agreed with R2A agar 377 times, for an accuracy rate of 92.6 percent. The authors recommend that clinicians consider using HPC Samplers, as they are a u ser-friendly, economical means of monitoring compliance of DUW quality .