M. Carta et al., DETERMINATION OF HBA(1C) IN THE PRESENCE OF HEMOGLOBIN-VARIANTS - COMPARISON OF 3 HPLC TECHNIQUES, European journal of clinical chemistry and clinical biochemistry, 35(12), 1997, pp. 923-925
The presence of Hb variants may cause analytical interference in HbA(1
c), values measured with the HPLC techniques currently used in Italian
laboratories. In this study performance of a new HPLC system, (HA-814
0, Menarini) was compared with two other traditional HPLC systems (HA-
8121, Menarini and Diamat, Bio-Rad). The HA-8140 system detects the Kb
variant possibly present in an independent peak. The integration area
of such a peak is not computed in the calculation of the percentage v
alue of HbA(1c). In this manner the underestimation of values obtained
with traditional HPLC systems is avoided.