A. Credi et L. Prodi, FROM OBSERVED TO CORRECTED LUMINESCENCE INTENSITY OF SOLUTION SYSTEMS- AN EASY-TO-APPLY CORRECTION METHOD FOR STANDARD SPECTROFLUOROMETERS, SPECT ACT A, 54(1), 1998, pp. 159-170
Citations number
23
Categorie Soggetti
Spectroscopy
Journal title
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY
Although spectrofluorimetry is a very sensitive analytical technique,
its use is limited by the non linear relationship between the concentr
ation of the analyte of interest and the generated electric signal. He
re we discuss a very easy correction method which takes into account a
ll the instrumental factors and transforms the observed intensity valu
e in a quantity, the corrected luminescence intensity, that is directl
y proportional to the concentration of the observed luminophore. At th
e end of the discussion, a general formula is presented together with
two examples illustrating the method and its application. (C) 1998 Els
evier Science B.V.