G. Halama et al., COMPOSITE X-RAY IMAGE ASSEMBLY FOR LARGE-FIELD DIGITAL MAMMOGRAPHY WITH ONE-DIMENSIONAL AND 2-DIMENSIONAL POSITIONING OF A FOCAL-PLANE ARRAY, Medical physics, 25(2), 1998, pp. 172-175
To demonstrate the feasibility of a novel large-field digital mammogra
phy technique, a 1024 x 1024 pixel Loral charge-coupled device (CCD) f
ocal plane array (FPA) was positioned in a mammographic field with one
-and two-dimensional scan sequences to obtain 950 x 1800 pixel and 360
0 x 3600 pixel composite images, respectively. These experiments verif
y that precise positioning of FPAs produced seamless composites and th
at the CCD mosaic concept has potential for high-resolution, large-fie
ld imaging. The proposed CCD mosaic concept resembles a checkerboard p
attern with spacing left between the CCDs for the driver and readout e
lectronics. To obtain a complete x-ray image, the mosaic must be repos
itioned four times, with an x-ray exposure at each position. To reduce
the patient dose, a lead shield with appropriately patterned holes is
placed between the x-ray source and the patient. The high-precision m
otorized translation stages and the fiber-coupled-scintillating-screen
-CCD sensor assembly were placed in the position usually occupied by t
he film cassette. Because of the high mechanical precision, seamless c
omposites were constructed from the subimages. This paper discusses th
e positioning, image alignment procedure, and composite image results.
The paper only addresses the formation of a seamless composite image
from subimages and will not consider the effects of the lead shield, m
ultiple CCDs, or the speed of motion. (C) 1998 American Association of
Physicists in Medicine.