MONITORING OF PESTICIDES IN RIVER WATER USING FULLY AUTOMATED ONLINE SOLID-PHASE EXTRACTION AND LIQUID-CHROMATOGRAPHY WITH DIODE-ARRAY DETECTION WITH A NOVEL FILTRATION DEVICE
S. Lacorte et al., MONITORING OF PESTICIDES IN RIVER WATER USING FULLY AUTOMATED ONLINE SOLID-PHASE EXTRACTION AND LIQUID-CHROMATOGRAPHY WITH DIODE-ARRAY DETECTION WITH A NOVEL FILTRATION DEVICE, Journal of chromatography, 795(1), 1998, pp. 71-82
Citations number
18
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A modification of the SAMOS (System for the Automated Monitoring of Or
ganic pollutants in Surface water) has been performed, which includes
the attachment of four refrigerated flasks prior to the preconcentrati
on unit, which act as sample reservoirs. River water is driven to the
SAMOS, where it is pumped inside the flasks and afterwards it is preco
ncentrated onto C-18 precolumns of the Prospekt (Spark Holland, Nether
lands), which is coupled on-line with liquid chromatography using the
HP1090 equipped with a diode array detector (Hewlett Packard, Waldbron
, Germany). The system is programmed in such a way that eight samples
are analyzed on each day and it can be operated unattended for at leas
t five days. In order to prevent clogging of the equipment due to the
presence of particulate matter, a novel filtering device, comprising a
50-cm stainless steel cylindrical tube with 10 mu m pores, connected
to a 190-mm diameter glass fiber filter with 1 mu m pores was installe
d. This configuration permitted satisfactory performance, with the una
ttended monitoring of Llobregat river water and with limits of detecti
on (LODs) of 30-100 ng/l. Terbuthylazine was detected in Llobregat riv
er effluent at a concentration of 13 mu g/l. Confirmatory analyses wer
e performed by liquid chromatography-atmospheric pressure chemical ion
ization-mass spectrometry. This latter technique was also used to scre
en organophosphorus pesticides in the Llobregat river. (C) 1998 Elsevi
er Science B.V.