PRACTICAL DESIGN OF A HIGH-STRENGTH BREAST GRADIENT COIL

Citation
Cf. Maier et al., PRACTICAL DESIGN OF A HIGH-STRENGTH BREAST GRADIENT COIL, Magnetic resonance in medicine, 39(3), 1998, pp. 392-401
Citations number
22
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
39
Issue
3
Year of publication
1998
Pages
392 - 401
Database
ISI
SICI code
0740-3194(1998)39:3<392:PDOAHB>2.0.ZU;2-D
Abstract
A high-strength three-axis local gradient coil set was constructed for MRI of the breast. Gradient fields with good uniformity (<10% deviati on from the desired gradient) over most of the volume required for bre ast imaging were generated with efficiencies of up to 3.3 mT/m/A. The coils will allow diffusion breast imaging in clinically acceptable exa mination times, The electrical design, water cooling system, and fabri cation techniques are described, Preliminary tests of the coil include d images of a grid phantom and diffusion measurements in a short-T-2 a garose gel phantom.