Rapid, simple and accurate chromatographic (HPLC) and spectrophotometr
ic methods for the determination of flecainide in tablets were elabora
ted. Methanol was found to be a suitable extraction solvent. The sampl
es were chromatographed on a LiChrosorb RP-18 column and the mobile ph
ase was acetonitrile/phosphate buffer pH 2.9/methanol (3:1:6, v/v/v).
Amiodarone was applied as an internal standard. The UV detection was c
arried out at 254 nm. The HPLC method was tested for linearity (over t
he range 4-20 mu g.ml(-1)). The recovery in HPLC was 99.27% (mean) and
in the UV-spectrophotometry 100.68% (mean).