Tj. Cohen et Dd. Lowenkron, THE EFFECTS OF PNEUMOTHORAX ON DEFIBRILLATION THRESHOLDS DURING PECTORAL IMPLANTATION OF AN ACTIVE CAN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR, PACE, 21(2), 1998, pp. 468-470
Pneumothorax has previously been reported 50 increase defibrillation t
hresholds and lead impedance in monophasic systems. This article demon
strates a case using an active can system in which Site lend impedance
between the right ventricular defibrillation coil and the device impr
oved from 70 Omega with pneumothorax to 48 Omega after resolution. The
defibrillation threshold also improved from > 30 J to less than or eq
ual to 10 J Of after the pneumothorax had resolved. We conclude that a
cute pneumothorax can increase the transthoracic lead impedance and de
fibrillation thresholds; however, with resolution both parameters can
return 50 acceptable levels with an active can system.