Rw. Sparidans et al., SEMIAUTOMATIC LIQUID-CHROMATOGRAPHIC ANALYSIS OF PAMIDRONATE IN SERUMAND CITRATE PLASMA AFTER DERIVATIZATION WITH 1-NAPHTHYLISOTHIOCYANATE, Journal of chromatography B. Biomedical sciences and applications, 705(2), 1998, pp. 331-339
Citations number
23
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
The semi-automatic method for the determination of the bisphosphonate
pamidronate in serum and citrate plasma involves a manual protein prec
ipitation with trichloroacetic acid and a manual coprecipitation of th
e bisphosphonate with calcium phosphate, followed by an automated soli
d-phase extraction on anion-exchange columns. After off-line evaporati
on of the extract under nitrogen and reconstitution in water, the auto
matic procedure is continued by automatic derivatization with 1-naphth
ylisothiocyanate, ion-pair liquid-liquid extraction and a treatment wi
th hydrogen peroxide, prior to analysis by ion-pair HPLC and fluoresce
nce detection at 285/390 nm. The intra-and inter-day precisions are 1.
3 and 7%, respectively, for a standard of 100 ng ml(-1) pamidronate in
serum; the average accuracy for this standard is 107%. The lower limi
t of quantification is 20 ng ml(-1) pamidronate in 1 ml of human serum
. (C) 1998 Elsevier Science B.V.