GAS-CHROMATOGRAPH INJECTION LINER FOR CONTINUOUS ANALYTE ADMISSION INTO A MASS-SPECTROMETER

Citation
Tn. Corso et al., GAS-CHROMATOGRAPH INJECTION LINER FOR CONTINUOUS ANALYTE ADMISSION INTO A MASS-SPECTROMETER, Analytical chemistry, 70(5), 1998, pp. 1030-1032
Citations number
3
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032700
Volume
70
Issue
5
Year of publication
1998
Pages
1030 - 1032
Database
ISI
SICI code
0003-2700(1998)70:5<1030:GILFCA>2.0.ZU;2-0
Abstract
An inexpensive modification to a gas chromatography injector liner is reported that facilitates continuous admission of analyte into a gas c hromatograph/mass spectrometer (GC/MS) for methods development. The MS methods development liner can be made by making simple modifications to commercially available liners and fits into standard injectors in p lace of the normal liners without any need to break vacuum in the MS. The injector temperature and gas now rates are adjusted to provide app ropriate analyte levels in the MS, which can be admitted under conditi ons identical with those of real analyses, including co-admission of c olumn bleed. The device is particularly useful for development of tand em MS methods in GC/MS/MS instruments, which are configured with the G C as the sole sample inlet.