The aim of the study was to determine whether any difference would occ
ur in peak systolic velocity (PSV) and end diastolic velocity (EDV) me
asurements of color Doppler ultrasonography (CDU) between proximal and
distal segments of the cavernous arteries obtained at different times
. Twenty-six cases have undergone CDU three times in weekly intervals.
PSV and EDV measurements of cavernous arteries were obtained both pro
ximally and distally. The results have shown that statistically there
was no difference between each measurement obtained on either sides at
three different tests. (ANOVA PSV F(2-50) = 0.63, EDV F(2-50) = 0.81)
. Four of the 26 cases had different PSV results and six cases had dif
ferent EDV results in three CDU tests. There was a statistical differe
nce in measurements between proximal and distal segments of the cavern
ous arteries. (t-test t = 2.53, P = 0.015). In conclusion, repetition
of CDU would not be necessary in majority of the cases. However invest
igators should always be aware and cautious about falsely abnormal res
ults of CDU. Proximal segments of the cavernous arteries should be the
site to obtain true CDU measurements.