THE WAVE-GUIDE MACH-ZENDER INTERFEROMETER AS ATRAZINE SENSOR

Citation
Ef. Schipper et al., THE WAVE-GUIDE MACH-ZENDER INTERFEROMETER AS ATRAZINE SENSOR, Analytical chemistry, 70(6), 1998, pp. 1192-1197
Citations number
21
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032700
Volume
70
Issue
6
Year of publication
1998
Pages
1192 - 1197
Database
ISI
SICI code
0003-2700(1998)70:6<1192:TWMIAA>2.0.ZU;2-O
Abstract
Immunoanalytical techniques used in combination with a highly sensitiv e Mach-Zender waveguide sensor give a device that is capable of fast o n-line monitoring of immunoreactions. The on-line monitoring is especi ally attractive when the sensor can be used as an environmental probe. It is demonstrated that low concentrations of atrazine can be measure d with a sensitivity around the EC limit of 100 ng/L. These measuremen ts are performed with an inhibition technique. In combination with ''a dynamic slope method'' one complete measurement is done within 10 min . Sensor measurements show the same sensitivity and errors as parallel ELISAs.