RESULTS OF BILATERAL PRESSURE COMPARISONS IN THE VACUUM REGION BETWEEN THE ETL AND THE IMGC

Citation
M. Hirata et al., RESULTS OF BILATERAL PRESSURE COMPARISONS IN THE VACUUM REGION BETWEEN THE ETL AND THE IMGC, Metrologia, 34(5), 1997, pp. 421-422
Citations number
5
Journal title
ISSN journal
00261394
Volume
34
Issue
5
Year of publication
1997
Pages
421 - 422
Database
ISI
SICI code
0026-1394(1997)34:5<421:ROBPCI>2.0.ZU;2-3
Abstract
Primary vacuum systems of the Electrotechnical Laboratory (ETL) and th e Istituto di Metrologia ''G. Colonnetti'' (IMGC) were compared in the pressure range 10(-4) Pa to 10 Pa by using spinning rotors as transfe r gauges. Several rotors were calibrated at both laboratories in an ar gon atmosphere. The results obtained are in good agreement and are wit hin the uncertainty values of both the laboratories.