COMPARISON OF JOSEPHSON ARRAY VOLTAGE STANDARDS USING ZENER REFERENCE-STANDARDS

Citation
Z. Gengru et al., COMPARISON OF JOSEPHSON ARRAY VOLTAGE STANDARDS USING ZENER REFERENCE-STANDARDS, Metrologia, 34(5), 1997, pp. 423-425
Citations number
4
Journal title
ISSN journal
00261394
Volume
34
Issue
5
Year of publication
1997
Pages
423 - 425
Database
ISI
SICI code
0026-1394(1997)34:5<423:COJAVS>2.0.ZU;2-4
Abstract
The Josephson array voltage standard at the Beijing Institute of Radio Metrology and Measurement (BIRMM) has been used in China to realize t he volt since 1992. In order to verify the accuracy and reliability of the voltage standard at the BIRMM, a comparison of the Josephson arra y voltage standards at 1,018 V and 10 V between the BIRMM, the Danish Institute of Fundamental Metrology (DFM) and the National Physical Lab oratory (NPL), UK, has been performed. Two Zener reference standards w ere used to transfer the voltage between the three laboratories. The r esults show a difference of less than 2x10(-8).