Jh. Christensen et al., RESPONSE CHARACTERISTICS AND APPLICATION OF CHALCOGENIDE GLASS CR(VI)SELECTIVE ELECTRODE, Sensors and actuators. B, Chemical, 45(3), 1997, pp. 239-243
A chalcogenide glass Cr(VI) selective electrode has been tested. The N
ernst plot is S-shaped, and the slope of the mid linear portion is abo
ut 90 mV decade(-1). An analytically useful potential change from abou
t 10(-6)-10(-3) M Cr(VI) results. The response is related to a surface
sensing mechanism, which is of redox nature. The electrode is general
ly sensitive towards compounds with high oxidising and reducing powers
but insensitive towards ions with poor oxidising or reducing capaciti
es. The response is slow and depends on the reduction potential of the
analyt compounds, the exposure time and the concentration level. Ther
efore a flow-Injection-Analyser-system (FIA) is developed for the dete
rmination of Cr(VI) in aqueous environments. For the reduction of resp
onse times thereby enhancing the performance of the electrode reducing
agents or electrochemical reduction between FIA injections were succe
ssfully employed. It is concluded, that to a large extend the FIA-syst
em can be optimised to aqueous environments with Cr(VI) concentrations
down to 10(-6) M. Special applications is under development for on-li
ne measurements of Cr(VI) for production control in the electroplating
industries. (C) 1997 Elsevier Science S.A. All rights reserved.