A. Safavi et Mr. Baezzat, FLUOROMETRIC KINETIC-FIA DETERMINATION OF AG(I) BASED ON ITS CATALYTIC EFFECT ON THE REDUCTION REACTION OF SAFRANINE WITH IODIDE, Microchemical journal, 58(2), 1998, pp. 138-143
A sensitive kinetic fluorimetric flow injection method has been develo
ped for the determination of Ag(I) over the range 100-2500 ng. The met
hod is based on the catalytic effect of Ag(I) on the reduction reactio
n of safranine by iodide. Method development includes optimization of
reagent concentration, pH, reaction temperature, and flow variables. T
he optimized flow injection manifold yielded a detection limit of 50 n
g of Ag(I). Chemical interference studies showed that the method also
has good selectivity. The method was applied to the determination of s
ilver in expired black and white photographic film. (C) 1998 Academic
Press.