THE ENERGY RESPONSE OF TLD-500K THERMOLUMINESCENCE DETECTORS OVER THERANGE OF 12-45 KEV UNDER SYNCHROTRON-RADIATION

Citation
Vs. Kortov et al., THE ENERGY RESPONSE OF TLD-500K THERMOLUMINESCENCE DETECTORS OVER THERANGE OF 12-45 KEV UNDER SYNCHROTRON-RADIATION, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 405(2-3), 1998, pp. 289-291
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
405
Issue
2-3
Year of publication
1998
Pages
289 - 291
Database
ISI
SICI code
0168-9002(1998)405:2-3<289:TEROTT>2.0.ZU;2-G
Abstract
The position of the temperature maximum, the geometrical factor and th e half-width of the main dosimetric peak of TLD-500K thermoluminescenc e detectors employing anion-defective corundum were studied as a funct ion of the dose under synchrotron radiation (SR) with an energy of 12. 6 keV. The dose curve was analyzed. The energy response of TLD-500K de tectors was examined over the SR range from 12 to 41 keV. It was found that the dependence has a maximum at 17 keV and sensitivity grows at energies over 30 keV.