STUDIES ON RAY-TRACING OF UV DETECTOR IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS

Authors
Citation
X. Xu et al., STUDIES ON RAY-TRACING OF UV DETECTOR IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS, Gaodeng xuexiao huaxue xuebao, 19(3), 1998, pp. 379-381
Citations number
4
Categorie Soggetti
Chemistry
ISSN journal
02510790
Volume
19
Issue
3
Year of publication
1998
Pages
379 - 381
Database
ISI
SICI code
0251-0790(1998)19:3<379:SOROUD>2.0.ZU;2-A
Abstract
A computer program is compiled to work out the ray-tracing of UV detec tor of high performance capillary Electrophoresis (HPCE). Based on cal culating results obtained by calculating, some practical methods for a djusting the distance between lens and capillary are provided to make UV detector work on the optimatic condition. The experiments with the aid of the method show a satisfactory detection sensitivity. The metho d for adjusting ray-tracing can be used into all kinds of HPCE UV-dete ctor.