X. Xu et al., STUDIES ON RAY-TRACING OF UV DETECTOR IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS, Gaodeng xuexiao huaxue xuebao, 19(3), 1998, pp. 379-381
A computer program is compiled to work out the ray-tracing of UV detec
tor of high performance capillary Electrophoresis (HPCE). Based on cal
culating results obtained by calculating, some practical methods for a
djusting the distance between lens and capillary are provided to make
UV detector work on the optimatic condition. The experiments with the
aid of the method show a satisfactory detection sensitivity. The metho
d for adjusting ray-tracing can be used into all kinds of HPCE UV-dete
ctor.