A sequential injection analysis (SIA) assembly for the atomic absorpti
on determination of Fe(III) in natural waters is proposed. Iron is pre
concentrated on a microcolumn packed with a chelating resin (Chelex 10
0) that is inserted in the manifold. The sample is passed through the
column and the iron retained by the resin is subsequently eluted with
2 M HNO3. The proposed SIA system affords automatic preconcentration,
elution, detection of Fe(III), data acquisition and treatment. When 9
ml of iron solution containing 0.4 or 1 mg l(-1) was passed through th
e resin, the retention efficiency was 93.1 +/- 0.6 and 7.4 +/- 3.0% re
spectively, and when 27 ml of iron solution of 0.2 mg l(-1) was precon
centrated, the retention was 8.4 +/- 2.9%. The detection limits thus a
chieved is 12 mu g l(-1) when 9 ml of sample are preconcentrated and 6
mu g l(-1) for 27 ml. (C) 1997 Elsevier Science B.V.