EVALUATION OF A FILLING SYSTEM FOR BINARY PEDIATRIC SOLUTIONS

Citation
D. Combeau et al., EVALUATION OF A FILLING SYSTEM FOR BINARY PEDIATRIC SOLUTIONS, Pharmacy world & science, 20(1), 1998, pp. 32-37
Citations number
12
Categorie Soggetti
Pharmacology & Pharmacy
Journal title
ISSN journal
09281231
Volume
20
Issue
1
Year of publication
1998
Pages
32 - 37
Database
ISI
SICI code
0928-1231(1998)20:1<32:EOAFSF>2.0.ZU;2-9
Abstract
The aim of this study was to assess the use of an automatic filling sy stem (Siframix(TM) M31 and M32 system) to prepare pediatric parenteral nutrition. volumetric accuracy was measured for each siframix(TM) sys tem loads cells (< 5% for 5 ml with the Siframix(TM) M32 and < 5% for 9 ml with the Siframix(TM) M31) with sterile water for injection, The minimal 20 ml of flushing sterile water of the common tubulure of the Siframix(TM) M32 (p=0.211 for 20 ml and p=0.75 for 500 ml), the use of viscous solutions (70 % dextrose) on the Siframix(TM) M31 (p=0.28 for 20 ml and p=0.12 for 500 ml) IX and the use of a special tubulure for using E.V.A. Luer-lock bags (p=0.89 for 20 mi and p=0.103 for 500 mi) do not modify the accuracy. Changing bags or bottles during the filli ng operation modify the accuracy (p=0.004 for 20 ml and p=0.009 for 50 0 ml). A flushing operation is necessary to lower the risk of electrol ytic pollution for the filling of little bags. The filling speed for e ach module as also measured (the maximal filling speed was five liters per minute). The Siframix(TM) system allows one to prepare pediatric parenteral nutrition bags when volumes are above 4 ml and with adapted source solutions in terms of concentration and conditioning volumes.