C. Haraldsson et al., RAPID, HIGH-PRECISION POTENTIOMETRIC TITRATION OF ALKALINITY IN OCEANAND SEDIMENT PORE WATERS, Deep-sea research. Part 1. Oceanographic research papers, 44(12), 1997, pp. 2031-2044
A system for rapid, high precision potentiometric determination of alk
alinity in sea water and sediment pore water is presented. Two titrati
on units were used: a 40 ml unit for seawater and a small volume unit
for sediment pore water. Titration time was normally less than 10 minu
tes per sample, including sample exchange. With a 40 ml sample volume,
the relative standard deviation of the alkalinity obtained in the lab
oratory was 0.05% and at sea 0.1%. The small-volume system (0.5-1.5 ml
) gave a precision of 0.07%. Five titration points, in two groups afte
r the second equivalence point; were used to evaluate the equivalence
volume. Results from equilibrium calculations and computer simulated a
lkalinity titrations show that it was possible to use a non-modified G
ran function [(V-0 + v)10((E/Z))] and still achieve good accuracy and
precision. (C) 1998 Elsevier Science Ltd. All rights reserved.