Z. Yu et D. Westerlund, CHARACTERIZATION OF THE PRECOLUMN BIOTRAP-500 C-18 FOR DIRECT-INJECTION OF PLASMA SAMPLES IN A COLUMN-SWITCHING SYSTEM, Chromatographia, 47(5-6), 1998, pp. 299-304
Citations number
24
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A new restricted access media (RAM) type of pre column, BioTrap 500 C-
18, for direct injection of plasma samples in column-switching systems
was evaluated with respect to the elution of plasma proteins in diffe
rent mobile phases, the loading capacity of plasma samples, the chroma
tographic behavior during plasma injections and protein contamination
of the packing and sealings. More than 95 % of plasma proteins could b
e, excluded from the precolumn within three minutes for all selected m
obile phases. Quantitative analyte recoveries could be obtained by inj
ecting plasma samples ranging from 5 to 500 mu L with the analyte mass
> 150 ng onto a BioTrap 500 C-18 column (20 x 4 mm I.D.). One precolu
mn tolerated about 15 mL of plasma injection without noticeable change
in retention and pressure. Clogging of the precolumn was encountered
(greater than or equal to 45 mL of plasma) due mainly to the adsorptio
n of proteins on the packing. The performance of the analytical column
(Kromasil C-18) was also examined. The column efficiency decreased by
60 % after processing 45 mL plasma in total.