A SIMPLE ADIABATIC LOW-TEMPERATURE CALORIMETER BASED ON A HELIUM REFRIGERATOR SYSTEM

Citation
U. Wolf et al., A SIMPLE ADIABATIC LOW-TEMPERATURE CALORIMETER BASED ON A HELIUM REFRIGERATOR SYSTEM, Thermochimica acta, 310(1-2), 1998, pp. 37-42
Citations number
7
Categorie Soggetti
Chemistry Analytical","Chemistry Physical
Journal title
ISSN journal
00406031
Volume
310
Issue
1-2
Year of publication
1998
Pages
37 - 42
Database
ISI
SICI code
0040-6031(1998)310:1-2<37:ASALCB>2.0.ZU;2-I
Abstract
The construction of an adiabatic calorimeter for the 15-300 K range is described. It is fully automated and set up in a helium refrigerator system. The operation of the calorimeter was tested by measuring the s pecific heat capacity of a pure copper sample. The results show good a greement with the standard literature values. The specific heat capaci ty of a magnesium hydride sample was also determined. For MgH2, the st andard molar entropy of S-o(298.15)=30.64 J mol(-1)K(-1) was calculate d from the obtained data. (C) 1998 Elsevier Science B.V.