O. Francis et al., CALIBRATION OF A SUPERCONDUCTING GRAVIMETER BY COMPARISON WITH AN ABSOLUTE GRAVIMETER FG5 IN BOULDER, Geophysical research letters, 25(7), 1998, pp. 1075-1078
This paper reports the results of a calibration of a superconducting g
ravimeter using simultaneous measurements of Earth tides with an absol
ute gravimeter. Nine consecutive days of observations were made with b
oth the C024 superconducting gravimeter and the FG5-202 absolute gravi
meter at the NOAA Table Mountain Gravity Observatory near Boulder, Col
orado. The precision of the calibration factor is better than 0.1%. Th
e calibration factor obtained in this fashion agrees well with that ob
tained from a moving platform. This experiment provides a noise estima
te for both the superconducting and absolute gravimeter. In addition,
the local air pressure admittances of both instruments compare well wi
th a value close to -0.35 mu Gal/mbar.