10-VOLT JOSEPHSON VOLTAGE STANDARD AT THE ETL
Citation
Y. Murayama et al., 10-VOLT JOSEPHSON VOLTAGE STANDARD AT THE ETL, IEEE transactions on instrumentation and measurement, 46(2), 1997, pp. 233-236
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
SICI code
0018-9456(1997)46:2<233:1JVSAT>2.0.ZU;2-X
Abstract
A voltage standard system using a 10 V Josephson junction array has be
en developed, The uncertainty is 0.6 x 10(-8) for measurement of the 1
0 V output of a Zener diode reference standard, A comparison of the 10
V array system with the conventional 10 V measurement system composed
of a 1 V array and a voltage divider was performed through measuremen
ts of a 10 V output of a Zener, The result agreed within the uncertain
ty (2 x 10(-8)) of the comparison, The 10 V output voltages of a Zener
reference standard are measured consecutively by the 10 V array syste
m.