ATTENUATION OF MICROWAVE FILTERS FOR SINGLE-ELECTRON TUNNELING EXPERIMENTS
Citation
A. Fukushima et al., ATTENUATION OF MICROWAVE FILTERS FOR SINGLE-ELECTRON TUNNELING EXPERIMENTS, IEEE transactions on instrumentation and measurement, 46(2), 1997, pp. 289-293
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
SICI code
0018-9456(1997)46:2<289:AOMFFS>2.0.ZU;2-Y
Abstract
Two types of microwave filters, metallic powder fitters and a tilter u
sing a resistive coaxial cable, were tested. Attenuation of the metall
ic powder filters using copper or stainless steel (SUS304L) was measur
ed at 300 K, 77 K, and 4.2 K, and it was found that SUS304L powder of
nominal 30 mu m grain size gives best result as the microwave filter f
or single electron tunneling (SET) experiments, Attenuation of a coaxi
al filter, or Thermocoax with SMA connectors attached at both ends is
larger at low temperatures than that at room temperatures, and the tem
perature dependence of the attenuation does not agree with Zorin's mod
el, The filter arrangement in our cryostat designed for SET experiment
s, is also reported.