C. Masquet et al., NORMAL LEFT-VENTRICULAR VOLUMES AND EJECTION FRACTION - ASSESSMENT WITH QUANTITATIVE DIGITAL CARDIOANGIOGRAPHY, International journal of cardiac imaging, 14(1), 1998, pp. 27-36
Citations number
16
Categorie Soggetti
Cardiac & Cardiovascular System","Radiology,Nuclear Medicine & Medical Imaging
The aim of this study was to assess the accuracy and precision of a di
gital angiography system, the General Electric DX system, as compared
to conventional left ventricular cineangiography. After a phantom stud
y, sixty patients with normal coronary arteries and normal left ventri
cles (LV) were studied. The LV contours were manually traced from raw
or subtracted digital runs and from the cine films in an independent m
anner. A regression formula was used to adjust the DX derived data to
the cine film results (V-corrected = 0.693 V-measured + 8.65 mi). Ther
e was an excellent correlation between the two methods in the end-dias
tolic volume index (I = 0.97, sd = 4.2 ml/m(2)), the end-systolic volu
me index (r = 0.95, sd = 2.7 ml/m(2)), the stroke volume index (r = 0.
93, sd = 4.1 ml/m(2)), and the ejection fraction (r = 0.85, sd = 3.45%
). We conclude that manual tracing of LV contours from either raw or s
ubtracted digital images provides reliable and accurate measurement of
LV volume and ejection fraction.