Abstract
A newly developed method allows dual six-post automatic network analyz ers (ANA's) to be calibrated with a single, known one-port termination instead of air line standards. This technique is especially useful fo r calibrations below 30 MHz, where air lines cannot be adequately char acterized, The choice of the value of the standard is also discussed.