P. Jaqueolmedo et al., NON SEGMENTED CONTINUOUS-FLOW SYSTEM WITH POLAROGRAPHIC DETECTION FORTHE DETERMINATION OF RANITIDINE IN PHARMACEUTICAL FORMULATIONS AND URINE, Boletin de la Sociedad Chilena de Quimica, 43(1), 1998, pp. 45-53
In this work we developed a method by flow injection analysis for the
determination of ranitidine in pharmaceutical forms and urine, using a
polarographic flow through cell as detection system which has been de
signed recently in our laboratory. The method is based on the reductio
n of the group nitro of the drug to the hydroxylamine derivative by ap
plying a potential of -1.40 V vs mercury pool electrode. It was found
a determination range of 2.38x10(-5) -1.0x10(-3) M ranitidine, with a
detection limit of 7.15x10(-6) M. The relative standard deviation (n =
11) for a concentration level of 1.0x10(-4) M was 2.1%, which is comp
arable with the obtained by the manual methodology. The accuracy of th
e method was evaluated by comparison of the results with those obtaine
d by d.c. polarography and UV spectrophotometry and the results were q
uite comparable.