DEVELOPMENT OF A QUANTITATIVE VANCOMYCIN IMMUNOASSAY FOR THE ABBOTT AXSYM ANALYZER

Citation
M. Adamczyk et al., DEVELOPMENT OF A QUANTITATIVE VANCOMYCIN IMMUNOASSAY FOR THE ABBOTT AXSYM ANALYZER, Therapeutic drug monitoring, 20(2), 1998, pp. 191-201
Citations number
38
Categorie Soggetti
Pharmacology & Pharmacy","Public, Environmental & Occupation Heath",Toxicology,Biology
Journal title
ISSN journal
01634356
Volume
20
Issue
2
Year of publication
1998
Pages
191 - 201
Database
ISI
SICI code
0163-4356(1998)20:2<191:DOAQVI>2.0.ZU;2-P
Abstract
A novel fluorescence polarization immunoassay for vancomycin on Abbott AxSYM analyzer is described. The immunoassay allows for the accurate quantification of vancomycin in the presence of the crystalline degrad ation product (CDP). It displays dilution linearity from 1.0 mu g/ml t o 100.0 mu g/ml, coefficients of variation ranging from 2.94% to 4.26% , recovery from 98% to 105%, and a sensitivity of <2.0 mu g/ml. The as say demonstrates no cross-reactivity to crystalline degradation produc t, and to commonly-prescribed and over-the-counter drugs, as well as a minimum interference from endogenous substances.