As. Manolis et al., SUCCESSFUL PERCUTANEOUS EXTRACTION OF PACEMAKER LEADS WITH A NOVEL (VASCOEXTOR) PACING LEAD REMOVAL SYSTEM, The American journal of cardiology, 81(7), 1998, pp. 935
A new pacing lead extraction system (VascoExtor, VascoMed, Germany) wi
th a universally applicable locking stylet was used over a period of 1
2 months to extract 25 permanent pacemaker leads (mean implantation ti
me 3.9 +/- 3.8 years) in 16 patients who had pacemaker infection (n =
13), lead dysfunction (n 2). or before an automatic defibrillator impl
ant (n = 1). With use of this system. we were able to successfully and
safely remove 24 of 25 chronic pacing leads (96%) in 15 of 16 patient
s (93%) with sole use of the locking stylet in 81%, and with use of an
array of ancillary tools in the remaining 19% of patients.