PERFORMANCE OF THE ATOM-235 SYRINGE INFUSION-PUMP UNDER HYPERBARIC CONDITIONS

Citation
Da. Story et al., PERFORMANCE OF THE ATOM-235 SYRINGE INFUSION-PUMP UNDER HYPERBARIC CONDITIONS, Anaesthesia and intensive care, 26(2), 1998, pp. 193-195
Citations number
7
Categorie Soggetti
Anesthesiology,"Emergency Medicine & Critical Care
ISSN journal
0310057X
Volume
26
Issue
2
Year of publication
1998
Pages
193 - 195
Database
ISI
SICI code
0310-057X(1998)26:2<193:POTASI>2.0.ZU;2-M
Abstract
Inotrope-dependent patients may become haemodynamically unstable durin g treatment in the hyperbaric chamber. One possible cause is altered p erformance of infusion devices. We tested six Atom 235 syringe infusio n pumps under hyperbaric conditions of 2.8 atmospheres absolute in a m ultiplace chamber using a gravimetric method. There was a small but si gnificant reduction in infused volume during a thirty-minute period in cluding compression time. We concluded that altered inotrope delivery may explain at least some of the haemodynamic instability sometimes se en, although a physiologic effect of hyperbaric treatment cannot be ex cluded.