Da. Story et al., PERFORMANCE OF THE ATOM-235 SYRINGE INFUSION-PUMP UNDER HYPERBARIC CONDITIONS, Anaesthesia and intensive care, 26(2), 1998, pp. 193-195
Citations number
7
Categorie Soggetti
Anesthesiology,"Emergency Medicine & Critical Care
Inotrope-dependent patients may become haemodynamically unstable durin
g treatment in the hyperbaric chamber. One possible cause is altered p
erformance of infusion devices. We tested six Atom 235 syringe infusio
n pumps under hyperbaric conditions of 2.8 atmospheres absolute in a m
ultiplace chamber using a gravimetric method. There was a small but si
gnificant reduction in infused volume during a thirty-minute period in
cluding compression time. We concluded that altered inotrope delivery
may explain at least some of the haemodynamic instability sometimes se
en, although a physiologic effect of hyperbaric treatment cannot be ex
cluded.