A METER FOR MEASURING HYDROGEN CONCENTRATION IN ARGON COVER GAS OF SODIUM CIRCUITS - DESIGN AND DEVELOPMENT

Citation
Kh. Mahendran et al., A METER FOR MEASURING HYDROGEN CONCENTRATION IN ARGON COVER GAS OF SODIUM CIRCUITS - DESIGN AND DEVELOPMENT, Industrial & engineering chemistry research, 37(4), 1998, pp. 1398-1403
Citations number
13
Categorie Soggetti
Engineering, Chemical
ISSN journal
08885885
Volume
37
Issue
4
Year of publication
1998
Pages
1398 - 1403
Database
ISI
SICI code
0888-5885(1998)37:4<1398:AMFMHC>2.0.ZU;2-7
Abstract
Monitoring of hydrogen in argon cover gas is one of the effective meth ods used to detect an event of steam/water leakage into sodium in the steam generators of a fast reactor during its startup and low-power op eration. A diffusion-based hydrogen meter employing a thermal conducti vity detector (TCD) and argon as carrier gas is a simple, convenient, and inexpensive device for this purpose. This paper describes the fund amental criteria involved in designing such meters and experimental re sults obtained during the calibration of two such meters.