3-DIMENSIONAL FAST SPIN-ECHO MR OF THE INNER-EAR - ULTRA-LONG ECHO TRAIN LENGTH AND HALF-FOURIER TECHNIQUE
Citation
S. Naganawa et al., 3-DIMENSIONAL FAST SPIN-ECHO MR OF THE INNER-EAR - ULTRA-LONG ECHO TRAIN LENGTH AND HALF-FOURIER TECHNIQUE, American journal of neuroradiology, 19(4), 1998, pp. 739-741
Categorie Soggetti
Clinical Neurology","Radiology,Nuclear Medicine & Medical Imaging
SICI code
0195-6108(1998)19:4<739:3FSMOT>2.0.ZU;2-Q
Abstract
We compared the image quality of the newly developed ultra-long echo t
rain length (ETL) 3-D fast spin-echo (FSE) and half-Fourier technique,
which is performed in less than 3 minutes, with the conventional 3D-F
SE imaging technique, which takes 15 minutes, in assessing MR examinat
ions of the inner ear. The new method's images were almost comparable
to the conventional 3D-FSE images in depicting anatomic details and pa
thologic findings. Implementation of the ultra-long ETL and half-Fouri
er 3D-FSE imaging technique enables acquisition of inner ear MR studie
s in a vastly reduced time and with high spatial resolution without si
gnificant penalty, opening the possibility for low-cost screening of a
coustic tumors without contrast enhancement in less than 3 minutes.