Although multiecho imaging may be used to measure transverse relaxatio
n (T-2), B-1 and B-0 inhomogeneity generally gives rise to unwanted co
herence pathway signals which result in erroneous T-2 measurements. On
e approach to suppressing this unwanted signal is to center each rf re
focusing pulse between spoiler gradients which dephase the unwanted si
gnal; however, hardware limitations often dictate the use of suboptima
l spoiler gradients, that is, gradients that cannot provide sufficient
dephasing strength. Using simulations, this work demonstrates that by
means of a small additional spoiler gradient prior to the first rf re
focusing pulse it is possible to reduce substantially the contribution
from unwanted coherence pathways in multiecho imaging studies that us
e suboptimal spoiler gradients. This reduction of unwanted signal resu
lts in measured T-2 values within approximate to 1% of values obtained
using spoiler gradients of optimal strength. These results were found
for a wide range of biologically relevant T-1 and T-2 values, missett
ings of the rf refocusing pulse as large as 5%, and frequency offsets
of up to 25 Hz. Multiecho image data agreed with the simulations. Usin
g the additional spoiler gradient it is possible to reduce spoiler gra
dient strengths by up to 75%. (C) 1998 academic Press.