EFFICIENCY CALIBRATION OF A LIQUID SCINTILLATION-COUNTER FOR Y-90 CHERENKOV COUNTING

Citation
F. Vaca et al., EFFICIENCY CALIBRATION OF A LIQUID SCINTILLATION-COUNTER FOR Y-90 CHERENKOV COUNTING, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 406(2), 1998, pp. 267-275
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
406
Issue
2
Year of publication
1998
Pages
267 - 275
Database
ISI
SICI code
0168-9002(1998)406:2<267:ECOALS>2.0.ZU;2-Q
Abstract
In this paper a complete and self-consistent method for Sr-90 determin ation in environmental samples is presented. It is based on the Cheren kov counting of Y-90 with a conventional liquid scintillation counter. The effects of color quenching on the counting efficiency and backgro und are carefully studied. A working curve is presented which allows t o quantify the correction in the counting efficiency depending on the color quenching strength. (C) 1998 Elsevier Science B.V. All rights re served.