Un. Andersen et al., IN-SERIES COMBINATION OF A MAGNETIC-SECTOR MASS-SPECTROMETER WITH A TIME-OF-FLIGHT QUADRATIC-FIELD ION MIRROR, Review of scientific instruments, 69(4), 1998, pp. 1650-1660
A tandem mass spectrometer consisting of a double-focusing magnetic-se
ctor mass analyzer in series with a time-of-flight (TOF) mass analyzer
has been designed and constructed. The TOF analyzer was a quadratic-f
ield ion mirror. The method of ionization used was matrix-assisted las
er desorption/ionization. Precursor ions were mass selected with the m
agnetic-sector analyzer, and time focused by ion bunching prior to fra
gmentation in a collision cell. The fragment ions were mass analyzed w
ith the TOF analyzer, which possessed the property that residence time
s (i.e., times of flight) in the mirror were independent of ion veloci
ty. The theoretical background to the instrumental design is presented
. Experimental results are presented, showing resolutions of 4000 in f
ragment ion spectra and demonstrating effective high-energy collision-
induced decomposition of peptide molecule ions. (C) 1998 American Inst
itute of Physics.