Pa. Cauchickmiguel et Tg. King, FACTORS WHICH INFLUENCE CMM TOUCH TRIGGER PROBE PERFORMANCE, International journal of machine tools & manufacture, 38(4), 1998, pp. 363-374
Since the introduction of Go-ordinate Measuring Machines (CMMs), probe
systems have been a matter of concern with regard to machine's error
sources. Initially CMMs used land sometimes still use) solid probes, b
ut for the last 20 years they have extensively employed touch trigger
probes. Using either probe type, CMM measurement results are affected
by probing errors. This paper presents some results from probe perform
ance verification, i.e. results from tests on a kinematic-resistive to
uch trigger probe to quantify its accuracy. The paper also emphasizes
sources of errors which affect probe performance as well as probe repe
atability. (C) 1998 Elsevier Science Ltd. All rights reserved.