A. Heinig et al., A NEW NEEDLE SYSTEM FOR PREOPERATIVE LOCA LIZATION AND CORE BIOPSIES OF SUSPICIOUS BREAST-LESIONS IDENTIFIED BY MAGNETIC-RESONANCE-IMAGING, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 166(4), 1997, pp. 342-345
Purpose: Examinations of a new needle system for localisation and core
biopsies of suspicious breast lesions, detected by MRI alone, were pe
rformed. It was investigated whether this new needle system could redu
ce and avoid artifacts. Material and methods: The new needle system wa
s examined by Spin Echo Sequence and a Fast Gradient Echo Sequence on
Magnetom Impact, 1.0 Tesla (in vitro-examinations). The sizes of the c
oaxial needle system on the MR images were examinated on various slice
thicknesses and with several wires in this needle system. Results: In
the Spin Echo Sequence the measured size of the coaxial needle is the
same as it's really size. In the Fast Gradient Echo Sequence the meas
ured size is about the double of the really size. Conclusion: The pres
ented coaxial system will allow improved visualisation for both preope
rative localisation and core biopsy.