AN AMPEROMETRIC DETECTOR WITH A TUBULAR ELECTRODE DEPOSITED ON THE CAPILLARY FOR CAPILLARY LIQUID-CHROMATOGRAPHY

Citation
T. Li et al., AN AMPEROMETRIC DETECTOR WITH A TUBULAR ELECTRODE DEPOSITED ON THE CAPILLARY FOR CAPILLARY LIQUID-CHROMATOGRAPHY, Analytica chimica acta, 360(1-3), 1998, pp. 53-59
Citations number
17
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
360
Issue
1-3
Year of publication
1998
Pages
53 - 59
Database
ISI
SICI code
0003-2670(1998)360:1-3<53:AADWAT>2.0.ZU;2-1
Abstract
A procedure is described for the preparation of a tubular electrode by chemical deposition of platinum at the end of a fused-silica capillar y. The properties of the electrode were tested under liquid chromatogr aphic conditions, demonstrating that both the static and the dynamic b ehaviour of the detection system satisfy the requirements of capillary chromatographic separations and compare well with a wall-jet amperome tric system and with UV photometric detection. The detection system de scribed is easy to prepare and does not require any time-consuming pos itioning of the electrode system as it is integrated into the separati on part of the apparatus. (C) 1998 Elsevier Science B.V.