T. Li et al., AN AMPEROMETRIC DETECTOR WITH A TUBULAR ELECTRODE DEPOSITED ON THE CAPILLARY FOR CAPILLARY LIQUID-CHROMATOGRAPHY, Analytica chimica acta, 360(1-3), 1998, pp. 53-59
A procedure is described for the preparation of a tubular electrode by
chemical deposition of platinum at the end of a fused-silica capillar
y. The properties of the electrode were tested under liquid chromatogr
aphic conditions, demonstrating that both the static and the dynamic b
ehaviour of the detection system satisfy the requirements of capillary
chromatographic separations and compare well with a wall-jet amperome
tric system and with UV photometric detection. The detection system de
scribed is easy to prepare and does not require any time-consuming pos
itioning of the electrode system as it is integrated into the separati
on part of the apparatus. (C) 1998 Elsevier Science B.V.