BIOMECHANICAL PROPERTIES OF ABSORBABLE IMPLANTS IN FINGER FRACTURES

Citation
F. Fitoussi et al., BIOMECHANICAL PROPERTIES OF ABSORBABLE IMPLANTS IN FINGER FRACTURES, Journal of hand surgery. British volume, 23B(1), 1998, pp. 79-83
Citations number
19
Categorie Soggetti
Orthopedics,Surgery
ISSN journal
02667681
Volume
23B
Issue
1
Year of publication
1998
Pages
79 - 83
Database
ISI
SICI code
0266-7681(1998)23B:1<79:BPOAII>2.0.ZU;2-1
Abstract
The mechanical rigidity of five different methods of pin fixation in t wo proximal phalangeal fracture models was studied and absorbable impl ants were compared with metallic implants in a biomechanical cadaver s tudy, Thirty phalanges were tested in apex palmar bending, compression and torsion. Results showed that rigidity of absorbable implants was comparable with metallic implants, except in torsion.