G. Bobe et al., SAMPLE PREPARATION AFFECTS SEPARATION OF WHEY PROTEINS BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, Journal of agricultural and food chemistry, 46(4), 1998, pp. 1321-1325
Measured concentrations of whey proteins in single milk samples often
differ by more than 20% when analyzed by different separation methods.
In the current study, we examined the effects of using guanidine hydr
ochloride or urea at zero and 8.5 h after sample preparation and the e
ffects of using dithiothreitol or a-mercaptoethanol on the separation
of bovine milk proteins by reversed-phase high-performance liquid chro
matography. Treatment with guanidine hydrochloride or urea resulted in
similar separation of milk proteins when samples were injected immedi
ately after preparation. Separation was repeatable over 8.5 h for samp
les prepared with guanidine hydrochloride, whereas the resolution of a
-lactalbumin and beta-lactoglobulin decreased for samples treated with
urea. Treatment with dithiothreitol improved the resolution of alpha-
lactalbumin and beta-lactoglobulin in comparison with treatment with 2
-mercaptoethanol. Quantitation of whey proteins is more reliable when
milk samples are treated with dithiothreitol and guanidine hydrochlori
de than when treated with 2-mercaptoethanol or urea.